Off 5.11.4.9 – Metrohm 774 Oven Sample Processor User Manual
Page 132

5 Detailed description
774 Oven Sample Processor, Instructions for use
126
.Keyboard
Disables all keys of the keyboard, except the
.Config
Locks the configuration menu
.Parameter
Locks the parameter menu
.Usermeth.Recall
Locks the function "recall method"
.UserMeth.Store
Locks the function "store method"
.Usermeth.Delete
Locks the function "delete method"
.Display
Disables the LCD display. The instrument will not sup-
port the display.
5.11.4.6
Setup.Mode.StartWait on,
off
Indefinite start delay. Only for remote control.
5.11.4.7
Setup.Mode.SendMeas.Sendstatus on,
off
Setup.Mode.SendMeas.Intervall 1…4…999
Automatic sending of the current measured values defined below at a se-
lectable interval (in seconds).
5.11.4.8
Setup.Mode.SendMeas.Meas.CyclNo on,
off
Setup.Mode.SendMeas.Meas.OvenTemp on,
off
Setup.Mode.SendMeas.Meas.GasFlow on,
off
Definition of the measured values to be sent automatically, see above.
5.11.4.9
Setup.AutoInfo.Status
on, off
Setup.AutoInfo.P on,
off
Setup.AutoInfo.Ch.G
on, off
Setup.AutoInfo.Ch.GC on,
off
Setup.AutoInfo.Ch.R on,
off
Setup.AutoInfo.Ch.S
on, off
Setup.AutoInfo.Ch.H
on, off
Setup.AutoInfo.Ch.C
on, off
Setup.AutoInfo.Ch.B
on, off
Setup.AutoInfo.Ch.F on,
off
Setup.AutoInfo.Ch.OM on,
off
Setup.AutoInfo.Ch.CM
on, off
Setup.AutoInfo.Heater.T
on, off
Setup.AutoInfo.Heater.S
on, off
Setup.AutoInfo.E on,
off
"on" means, the instrument automatically sends a corresponding message
when the specified event occurs.
.Status
Enables/disables the preset AutoInfo messages
.P
PowerOn: Simulation of PowerOn (see 5.11.4.10). Not from
mains.
Messages from changer functions:
.Ch.G
Go: Method has been started
.Ch.GC
Go Command: Start command has been received
.Ch.R
Ready: Status 'Ready' reached
.Ch.S
Stop: Status 'Stop' reached
.Ch.H
Hold: Status 'Hold' reached
.Ch.C
Continue: Resuming after Hold
.Ch.B
Begin: Begin of sample sequence
.Ch.F
Final: End of sample sequence
.Ch.OM
"Opening Moves": Begin of start sequence
.Ch.CM
"Closing Moves": Begin of final sequence
.Heater.T
Target: Change of target temperatur