5.3 Menu organization

774 Oven Sample Processor, Instructions for use

63

CONFIG

5.3.1 Configuration

Main Menu:

configuration

>auxiliaries

Open submenu with

configuration

>oven settings

Use <

Ï

> or <Ð> to move up or down one menu item

configuration

>rack definitions

Use or to move to first or last menu
item respectively

configuration

>dosing units

Return to normal state with

configuration

>RS232 settings




configuration

>auxiliaries

Basic Settings Submenu

Open the submenu with

>auxiliaries

dialog: english

Choice of dialog language

english, deutsch,

français,

español

>auxiliaries

display contrast 3

Setting display contrast

0…3…7

0 = no contrast

7 = large contrast

>auxiliaries

beeper: ON

Turn acoustic warning signal on or off

ON,OFF

>

auxiliaries

device label ********

Instrument label

8 ASCII characters

>auxiliaries

program 5.774.0010

Program version

read

only

The following three entries become effective after a RESET or switching the
changer on again.

>

auxiliaries

max. lift way 90

mm

Max. stroke path for lift

0…90…100 mm

This setting of the max. lift path is important as regards safety. Correctly en-
tering this value can prevent the puncture needle breaking off since the lift
cannot be moved lower than as far as the specified position.

use

to access next

higher level

use

to access next

higher level