Metrohm 774 Oven Sample Processor User Manual
Page 123

5.11 Description of the remote control commands
774 Oven Sample Processor, Instructions for use
117
5.11.1.7
Mode.SampleSeq.1.∗
.Move…, .Lift…, .Sample…,
etc. up to .100
.Heater…,.Def…, .Flow…, Dos…,
Scan…, .Ctrl…, .Wait…, .Rack, .Nop
Indexed sample sequence; its commands will be executed line by line in
processing. See start sequence (5.11.1.5).
5.11.1.8
Mode.FinalSeq.1.Cmd
NOP, MOVE, LIFT, STIR,
etc. up to .100
DEF, PUMP, DOS, SCAN,
SAMPLE, CTRL, WAIT, ENDSEQ
Defines the command of the indexed command line in a final sequence.
See start sequence (5.11.1.4).
5.11.1.9
Mode.FinalSeq.1.∗
.Move…, .Lift…, .Sample…,
etc. up to .100
.Heater…,.Def…, .Flow…, Dos…,
Scan…, .Ctrl…, .Wait…, .Rack, .Nop
Indexed final sequence; its commands will be executed line by line in proc-
essing. See start sequence (5.11.1.5).
5.11.1.10
Mode.Report.Assign
none, short, full, param,
config,
Usermeth
Mode.Report.TDelta 1…10…60
Report definition.
Only one report is possible. In the case of a 'full' result report, a measuring
point list with the oven temperatures as a function of time is output by con-
trast with the 'short' result report. The time interval (in seconds) between the
measuring points defines the node .TDelta. 400 measuring points are pos-
sible, see 5.11.3.2.
5.11.1.11
Mode.Changer.RackNo
0…16
Mode.Changer.L1Rate 3...12 mm/s
Mode.Changer.L2Rate 3...12 mm/s
Mode.Changer.ShRate 3...20 w/s
Mode.Changer.ShDir +,
–,
auto.
Mode.Changer.ModeSample
MOVE, display
Changer settings.
RackNo:
Rack number, forces the use of the defined rack with the
current method (0 = any rack).
L1Rate:
Lift speed for tower 1, in mm/sec
L2Rate:
Lift speed for tower 2, in mm/sec
ShRate:
Turning speed of the rack, in angular degrees/sec
ShDir:
Turning direction of the rack (ascending or descending
rack positions; auto. means automatic choice of the
shortest path)
ModeSample:
Reaction on missing sample beaker. (MOVE = next sam-
ple beaker will be chosen regarding the recent SAMPLE
command, display = display warning.)
5.11.1.12
Mode.TimeoutSet.HTime
off, 0...999
Mode.TimeoutSet.HAction cont,
STOP
Mode.TimeoutSet.STime
off, 0...999
Mode.TimeoutSet.SAction cont,
ERROR
Timeout settings. Timeout times are defined in seconds.
The HEATER timeout occurs if the set oven temperature (in the case of the
HEATER command) is not reached after the specified heating (in minutes).