beautypg.com

Bird Technologies BDS-Manual User Manual

Page 88

background image

74

Table 5-52 Standard Mode Instance Attributes

Table 5-53 Standard Mode Common Services

153

Get/
Set

NV

Is
Calibration
Disabled

BOOL

Option to turn off
measurement calibration.

FALSE for
calibrated
data (default),
TRUE for raw
data
(diagnostic).
Default:
FALSE

Hidden
attribute for
diagnostic
purposes only!

Attrib

Rule

NV

Name

Data

Type

Description

Semantics

of Value

Attrib Rule NV

Name

Data

Type

Description

Semantics

of Value

1

Get/
Set

NV

Fundamental
Selection

STRUCT
of:
Fundame
ntal

Fundamental frequency
and harmonic
component selection.

See the
Fundamenta
l structure.

2

Get

V

Num
Measurements

USINT

The number of voltage
and current harmonic
component
measurements
contained in the
following array.

Value range:
0-16

3

Get

V

Measurements

Array of:
STRUCT
of:
VIMeas

Voltage and current
harmonic component
measurements.

See VIMeas
structure.

4

Get

V

Num Intermods

USINT

The number of
intermodulation products
contained in the
following array.

Value range:
0-6

5

Get

V

Intermods

Array of:
STRUCT
of:
VIMeas

Voltage and current
intermodulation product
measurements.

See VIMeas
structure.

Service

Code

Class Instance

Service Name

Description of Service

14 (0x0E)

Yes

Yes

Get_Attribute_Single Returns the contents of the

specified attribute.

16 (0x10)

Yes

Yes

Set_Attribute_Single

Modifies an attribute.