39

Code

Description

Action

OOOO

No power applied.

Reboot the receiver by
disconnecting power and then
reconnecting power.

RRRR

CPU exception to unknown vector.

OROR

RAM Failure

ORRO

Bootloader checksum failure.

ORRR

Kernel failure.

OOOR

CPU Fault – Illegal instruction.

OORO

CPU Fault – Address error.

OORR

CPU Fault – Unimplemented A-line.

OROO

CPU Fault – Unimplemented F-line.

YYOO

Bootloader RAM test pass.

These LED codes are normally
displayed during receiver boot-up.
If the BDS reports any one of
these codes for an indefinite
amount of time reboot the
receiver by disconnecting power
and then reconnecting power.

YYYO

Bootloader loaded.

YYYY

Bootloader running.

GYYY

Operating system loaded.

GGYY

Operating system running.

GGGY

Main applications running, but calibration
data has not been loaded (the probe is
disconnected).

Attach the probe to the receiver
and wait approximately 1 minute.

GGGG

System is ready with all components
connected and all calibration data loaded.

No action. Ready for use.

YGGY

Loading FPGA image.

This code is normally displayed
during receiver boot-up. Update
the firmware if the BDS reports
this code for an indefinite
amount of time.

RYYY

DSP error – general or initialization failure.

Update the firmware.

YRYY

DSP error – DCM lock failure.

Reboot the receiver by
disconnecting power and then
reconnecting power.

YYRY

DSP error – PLL lock failure.

YRRY

DSP error – heartbeat flat-line. DSP stopped
producing data.

YYYR

Fatal application software error –
unrecoverable exception caught.

Update the firmware and/or
reboot the receiver by
disconnecting power and then
reconnecting power.

RYRY

Fatal application software error – assertion
failure.

YRYR

Fatal application software error –
segmentation fault.

YYRR

Fatal application software error – bus error.