Status leds – Bird Technologies BDS-Manual User Manual
Page 27

13
Status LEDs
The receiver has four status LEDs. Each LED can have four conditions, off (O), on green (G), on yellow (Y), and on red
(R). The four letters in the LED code column are arranged left-to-right as viewed on the instrument. The letters indi-
cate the condition of each LED. The meaning of the code is in the description column.
Code
Description
OOOO
No power applied
RRRR
CPU Exception to unknown vector
OROR
RAM Failure
ORRO
Bootloader checksum failure
ORRR
Kernel Failure
OOOR
CPU Fault - Illegal instruction
OORO
CPU Fault - Address error
OORR
CPU Fault - Unimplemented A-line
OROO
CPU Fault - Unimplemented F-line
YYOO
Bootloader RAM test pass
YYYO
Bootloader loaded
YYYY
Bootloader running
GYYY
Operating system loaded
GGYY
Operating system running
GGGY
Main applications running, but calibration data has not
been loaded (the probe is disconnected).
GGGG
System is ready with all components connected and
all calibration data loaded.
YGGY
Loading FPGA image.
RYYY
DSP Error – general or initialization failure.
YRYY
DSP error – DCM lock failure.
YYRY
DSP error – PLL lock failure.
YRRY
DSP error – heartbeat flat-line. DSP stopped
producing data.
YYYR
Fatal application software error – unrecoverable
exception caught.
RYRY
Fatal application software error – assertion failure.
YRYR
Fatal application software error – segmentation fault.
YYRR
Fatal application software error – bus error.