Status LEDs

The receiver has four status LEDs. Each LED can have four conditions, off (O), on green (G), on yellow (Y), and on red
(R). The four letters in the LED code column are arranged left-to-right as viewed on the instrument. The letters indi-
cate the condition of each LED. The meaning of the code is in the description column.

Code

Description

OOOO

No power applied

RRRR

CPU Exception to unknown vector

OROR

RAM Failure

ORRO

Bootloader checksum failure

ORRR

Kernel Failure

OOOR

CPU Fault - Illegal instruction

OORO

CPU Fault - Address error

OORR

CPU Fault - Unimplemented A-line

OROO

CPU Fault - Unimplemented F-line

YYOO

Bootloader RAM test pass

YYYO

Bootloader loaded

YYYY

Bootloader running

GYYY

Operating system loaded

GGYY

Operating system running

GGGY

Main applications running, but calibration data has not
been loaded (the probe is disconnected).

GGGG

System is ready with all components connected and
all calibration data loaded.

YGGY

Loading FPGA image.

RYYY

DSP Error – general or initialization failure.

YRYY

DSP error – DCM lock failure.

YYRY

DSP error – PLL lock failure.

YRRY

DSP error – heartbeat flat-line. DSP stopped
producing data.

YYYR

Fatal application software error – unrecoverable
exception caught.

RYRY

Fatal application software error – assertion failure.

YRYR

Fatal application software error – segmentation fault.

YYRR

Fatal application software error – bus error.