beautypg.com

Averaging, Figure 22 averaging dialog, Smoothing – Bird Technologies BDS-Manual User Manual

Page 49: Figure 23 smoothing dialog, Averaging smoothing

background image

35

Default Settings -

Clicking this button will restore the default sweep settings.

Resolution: 12-bit

Frequency Step Size: Uniform

No. Points: 2100 (Fixed)

Start Frequency: 1 MHz

Stop Frequency: 100 MHz

Frequency Step: 47.165 kHz

Filter resolution equals frequency step: Checked

Continuous Sweep: Checked

Averaging

Controls the number of sweeps to average. Sweeps are averaged successively as they are made available, so the sig-
nal-to-noise ratio improves over time. Up to 100 sweeps can be averaged. Sweep data averaging is only available
when the “Continuous Sweep” option is checked.

Figure 22 Averaging Dialog

Smoothing

Enabling smoothing will filter out noise by averaging a specified number of neighboring points.

Figure 23 Smoothing Dialog

Smoothing Window Size -

Determines how many points to average before and after a center point.

Smoothing Passes -

Takes multiple passes thought the sweep data, applying the same smoothing filter on each pass.

Note: Smoothing can be used in combination with data averaging. In can also be used on a single sweep
(“Continuous Sweep” can be enabled or disabled) or even be changed while sweeping.