Time Waveform Measurement Definitions
The Emonitor software and the Dynamic 2500 data collector support time
waveform measurement definitions. This topic lists the unique characteristics of
the Dynamix that you may need to know when setting up and collecting time
waveform data.
The Dynamix 2500 data collector cannot collect phase with time waveform
measurement definitions. You can choose to have the instrument collect a time
waveform only if a magnitude measurement at the same location is in alarm.
The Dynamic 2500 data collector supports integrated time waveform
measurements (single integration). This means you can collect a time waveform
with velocity units using an accelerometer or displacement units using a velocity
transducer.
The length of the time waveform depends on the maximum frequency you select
in the collection specification. You can calculate the number of time samples from
the number of spectral lines (Collection Specification dialog box) using this
formula.
You can calculate the total time required to collect a time waveform measurement
from the Number of lines value and the Maximum frequency value in the
Collection Specification dialog box (Setup > Collection Specification).
Table 32 - Collection specifications for spectrum measurement definitions
Transducer
Window
Signal Detection
Fmax
Lines
Phase
Order Norm?
Averages
Accelerometer
Hanning
Peak
2 kHz
400
No
No
4 linear
Accelerometer
Hanning
Peak
10 orders
400
No
Yes
4 linear
Accelerometer
Hanning
Peak
2 kHz
400
No
No
4 time synch.
Table 33 - Transducer specification in spectrum collection specifications
Name
Base Unit
Input Type
Units
Calibration Value
DC Offset
Accelerometer
Acceleration
ICP Accel
g’s
100
0
TIP
Set the Collect and store condition in the Edit Storage Specification
dialog box to On Magnitude Alarm for the storage specification for the
time waveform measurement definition.
time samples
spectral lines 2.56
=