Chapter 1 The Dynamix 2500 Data Collector
Status Indicators
The Dynamix 2500 data collector has four status indicators located in the
upper-right side of the instrument nameplate.
Figure 4 - Status Indicators
Table 7 - Status Indicators
Status
Indicator
Operation
State
Definition
Dual State
Status
Red
Charging Battery
Flashing
When first connected to the power adapter, the
Dynamix 2500 Data Collector tests the condition of
the battery pack. Within 30 seconds, the status
indicators should go to a solid state.
If the status indicator continues to flash, the
instrument has diagnosed a fault in the battery pack
and it will not charge the battery. The anomaly may
be that the cell temperature is too high or there is a
fault with the battery pack.
Solid
The battery pack is OK. The power adapter is
connected to the instrument and charging. The
battery pack is maintained by a trickle charge from
the power adapter for as long as it remains attached
to the instrument.
Data Collection
Solid
• An alarm condition has been set.
• Input signal is out of range.
• ICP check has failed and there is an ICP fault
condition.
• The power adapter is connected to the
instrument.
Blue
Communication
Solid
• Communicating to Emonitor through the USB
connection.
Amber
Data Collection
Solid
Acquisition sub-system is settling.
Green
Data Collection
Solid
Input signal is stable and data is ready to be stored.
32146-M
Dual Status: Red and Blue
Green
Amber