Care and use manual – Waters IC-Pak Column and Guard User Manual
Page 8

[ Care and Use ManUal ]
IC-Pak Column and Guard Column
8
5. Fill the flask to the mark with 18 megohm water. Tables 8 and 9
contain the species concentrations in the monovalent and divalent
cation working standards.
Table 8: Monovalent Cation Working Standard Concentrations
Table 9: Divalent Cation Working Standard Concentrations
e. Running the Calibration Standard and Determining
Column Efficiency
Refer to Table 10 for the parameters to run the calibration standard.
Table 10: Calibration Standard Parameters
* By half-height method on a low dispersion Waters ion
chromatograph
** Do not exceed a flow rate of 1.0 mL/min for the IC-Pak A HR
column. Flow rates higher than this may damage the column and void
the warranty.
f. Calculating Efficiency
Refer to Figure 6 for the appropriate peak to measure for your anion
or cation column. Use the equation in the figure below to compute
column efficiency for comparison to the value in the table.
Figure 6: Half Height Method Test Calculations
Monovalent Cation
Stock Solution
ppm
Lithium
0.1
Sodium
0.5
Ammonium
1.0
Potassium
1.0
Divalent Cation
Stock Solution
ppm
Magnesium
2
Calcium
4
Strontium
6
Barium
16
Column
Name
Optimal Flow
Rate (mL/min)
Test Sample
Efficiency
(Plates)*
Peak for Plate
Measurement
IC-Pak A
1.2
Anion
800
sulfate
IC-Pak C
1.2
Cation
1100
sodium
IC-Pak A HC
2.0
Anion
1500
sulfate
IC-Pak A HR
1.0**
Anion
2500
sulfate
IN
JE
C
T
W½
½h
h
= Column Efficiency (plates)
= Volume to peak apex (µL)
= Volume of peak at ½ peak height (µL)
N
VR
W ½
N = 5.54 x
2