[ CARE AND USE MANUAL ]

HILIC Quality Control Reference Material

CONT ENTS

I. INTRODUCTION

II. STORAGE AND STABILITY

III. EXAMPLE OF USING THE HILIC QCRM

IV. ORDERING INFORMATION

I. INTRODUCTION

The Quality Control Reference Material (QCRM) portfolio is a unique
collection of standards and mixtures. These products allow the user to
evaluate and benchmark their chromatography system before analysis
of critical material. The products in the portfolio are all specifically
formulated based on the expertise of Waters scientist.

The retention of polar analytes and metabolites is a chromatographic
challenge for scientists. Polar compounds often elute in the void of most
reversed-phase columns or end up co-eluting with endogenous species
present within various sample matrices in RPLC. Hydrophilic interaction
chromatography (HILIC) is a complementary technique to reversed phase
and normal phase, that can be used to successfully improve the retention
of very polar species.

1

The HILIC QCRM was designed as a standard to help benchmark this type
of separation. It is a four component mix containing an un-retained void
marker, a polar neutral compound, and two polar basic compounds that
provides a stable chromatographic separation for HILIC columns.

HILIC QCRM Components

Compound Name

Property

Concentration in Vial (mg/mL)

Acenaphthene

V0 marker

0.0190

Thymine

Polar neutral

0.0037

Adenine

Polar basic

0.0037

Cytosine

Polar basic

0.0077

Figure 1. Individual Compound Formulas.

acenaphthene

cytosine

thymine

adenine