Care and use manual – Waters IC-Pak Column and Guard User Manual
Page 6

[ Care and Use ManUal ]
IC-Pak Column and Guard Column
6
Temperature:
* Do not exceed a flow rate of 1.0 mL/min for the IC-Pak A HR 
column. Flow rates higher than this may damage the column and will 
void the warranty. 
General Considerations:
•
Sodium	silicate	leaching	from	glass	vials	will	cause	artifacts	when	
analyzing anions or cations. Plastic containers should be used for all 
solutions. 
•
Dedicate	columns	to	one	eluent	only,	if	possible.	Multiple	eluent	
changes may result in shortened column life. Dedicated columns should 
be used for either monovalent or divalent cation analysis.
•
Filter all aqueous buffers. Do not use turbid or cloudy buffers.
•
Remove	aliquots	from	the	sample	container	for	pH	readings,	and	then	
dispose of the contaminated aliquots.
•
Use a pH meter to measure the pH of eluents.
•
Protect	columns	from	vibration,	mechanical	shock,	and	rapid	changes	
in pressure, flow rate or solvent composition. Any thermal, physical or 
chemical shock (such as changing solvents rapidly or at high flow rates) 
can cause a loss of efficiency.
•
When	using	water,	treat	with	a	Milli-Q	water	system	capable	of	deliver-
ing 18 megohm water. Neither deionized water nor HPLC grade bottled 
water is acceptable because each may contain organic compounds 
which alter column selectivity.
•
DO	NOT	inject	concentrated	samples	directly	into	the	eluent.	Direct	
injection may cause precipitation of the salts in the sample. Dissolve 
(or dilute) samples in an appropriate volume of the eluent first. If other 
solvents must be used, be sure no precipitation occurs upon injection 
into the eluent. Always filter samples before use.
•
Highly	concentrated	samples	(greater	than	100	ppm	per	ion)	may	yield	
poor peak shape due to overloading the column. Dilute the sample 
before injection. When analyzing an unknown, prepare a 1:100 dilution 
as a first step in optimizing the method.
b. Efficiency Testing
Waters columns are tested for adherence to Waters specifications. 
Slight variations in results will occur depending on:
•	 Equipment	used
•	 Test	sample	makeup
•	 Equipment	settings	and	conditions
Perform an initial efficiency test before attempting the first analysis. 
Run the test sample using the calibration standards detailed in the 
following pages and record the results (retention time and the set-
tings used).
The initial efficiency test is performed by:
1. Preparing the eluent (Section III, a)
2. Preparing the calibration standards (Section IV, b to Section IV, d)
3. Running the calibration standard and determining column efficiency
(Section IV, e)
If problems occur during normal operation of the column, repeat the 
conditions for the initial efficiency test and compare the results. Dif-
ferences in the results may indicate the source of the problem.
c. Anion Calibration Standard Preparation
1. To prepare individual 1000 ppm (mg/L) stock standards, refer to Table
4 for the weight of salt required. Select the highest purity salt available, 
weigh the specified amount, and add to a 1 liter volumetric flask.
Table 4: Salt Weight for Anion Stock Solution Preparation
Anion (expressed as
compound listed)
Compound
Weight in grams
F
-
NaF
2.2101
CI
-
NaCl
1.6485
Br
-
NaBr
1.2877
NO
3
-
NaNO
3
1.3708
HP0
4
-2
KH
2
PO
4
1.4179
SO
4
-2
Na
2
SO
4
1.4787
Normal operation
25 °C
Limits
10 °C - 50 °C
