beautypg.com

Care and use manual – Waters Advanced Purification Glass Columns User Manual

Page 8

background image

[ Care and Use ManUal ]

Waters Advanced Purification Columns

8

Replacing an inlet connector O-ring

To replace the inlet connector O-ring:

1. Deactivate the inlet O-ring by turning the seal adjustment knob.

Remove the inlet connector assembly from the column by unscrewing
the plunger.

2. Use a small blunt tool, such as a push pin, to gently pry up the O-ring

so that you may roll it off the connector with your fingers. Carefully
push a new O-ring onto the connector tip. Be sure to select the proper
inlet connector O-ring.

Do not scratch the O-ring groove. If the groove is scratched, the column
will leak.

3. Return the inlet connector assembly to the column. Reactivate the

O-ring.

Replacing a filter O-ring

This small O-ring is located at the edge of the filter skirt. With the filter off,
roll the O-ring off the connector tip and replace it with a new one.

f. Storing the Column

Storage considerations

Store a packed glass column according to the recommendations of the
manufacturer of the packing material. Note the following:

Store the column with end plugs in place to prevent column drying.

Use bacteriostatic solutions in appropriate concentrations, for example,
0.2% sodium azide or 30% ethanol.

Remove all salts from the column before adding an organic solvent.
An organic solvent added to a salt can cause precipitation, resulting in
increased backpressure.

Extended storage in high salt buffers may result in the formation of
microcrystals which can lead to high backpressure. (Dissolve microcrys-
tals by slowly introducing a low ionic strength buffer).

Do not allow the mobile phase to freeze. Storage at 4 °C is acceptable.

IV. sCalIng up separatIons

a. Scaling a Separation

When scaling an analytical separation to a preparative separation, use the
following isolation approach:

1. Define the objective by determining the:

• Complexity of the sample mixture
• Component to be isolated
• Required quantity
• Degree of purity
• Properties of the components in the mixture

2. Perform crude separations to remove large quantities of extraneous

material or to isolate groups or classes of compounds. Simplify a com-
plex sample through coarse separation techniques, such as liquid-liquid
or solid-liquid extraction, crystallization, or precipitation. Centrifuge or
filter the sample to remove particulates.

3. Develop the separation method. To optimize the separation for com-

pounds of interest, make small-scale injections of the sample mixture,
varying the gradient.

4. Perform a loading study. Make progressively larger injections of the

sample mixture on a small scale column, such as the AP-Minicolumn,
to determine the effect of overload on resolution. This study defines the
amount of sample that can be separated in a column of a given volume.

5. Scale up the separation. Translate the small scale parameters to the

large scale system values. Adjust the:

Sample load

Flow rate

Gradient volume

b. Adjusting Sample Load and Flow Rate

When scaling the separation, use analytical and preparative
columns that are:

The same length

Packed with the same particle size material