beautypg.com

Care and use manual – Waters Advanced Purification Glass Columns User Manual

Page 4

background image

[ Care and Use ManUal ]

Waters Advanced Purification Columns

4

Preparing slurry solutions

To make slurry solutions, suspend one part of packing material
in three parts of degassed solution. The packing funnel can hold
40 mL of slurry.

Slurry-packing

To slurry-pack the column (Figure 4):
1. Unscrew the inlet connector from the inlet cap.

2. Screw the packing funnel into the inlet cap.

3. Clamp the column upright to a rigid support. Drain the outlet tubing to

waste.

4. Fill the column with the slurry. If the slurry does not flow into the

column, insert a pipette into the tip of the funnel to break the surface
tension.

5. When all of the packing material is in the column, plug the outlet

tubing.

6. Let the packing material settle. Remove excess solvent from the top of

the bed with a pipette. Do not disturb the bed.

7. Remove the funnel, and reattach the inlet connector.

8. Adjust the inlet connector until the plunger contacts the packed bed.

Ensure that the packed bed does not exceed the nominal column
length.

Figure 4: Slurry-Packing the Column

Dry-Packing the Column

Amount of packing material required
To determine the amount of packing material required, multiply the
column volume, V (mL), by the bulk density of the packing material,
d (g/mL):

Amount of packing material required = V x d

Waters Protein-Pak HR must be slurry-packed (see “Slurry-packing the
column”)

Determining density

To determine the ulk density of the packing material, put 19 g of
packing material into a 10 mL graduated cylinder. Tap the graduated
cylinder on a flat surface until the material is completely settled.
Note the volume and calculate the density.

Dry-packing

To dry-pack the column:

1. Unscrew the fine adjustment knob from the inlet cap.

2. Screw the packing funnel into the inlet cap.

3. Remove the outlet tubing.

4. Holding the column vertically, gradually fill the column with packing

material. While filling, tap the column lightly on a flat surface until the
material settles.

5. Remove the funnel and reattach the inlet connector and the outlet

tubing.

6. Adjust the inlet connector until the inlet plunger contacts the packed

bed.

Ensure that the packed bed does not exceed the nominal column
length.

Packing Material

Solution

Ion-exchange chromatography

High concentration buffer

Reversed-phase chromatography

Methanol

Hydrophobic-interaction
chromatography (HIC)

Water or low concentration buffer

Slurry

Funnel

Inlet cap

To waste

Packing Material

Bulk Density of Packing Material

Waters AccellPlus silica-based
packing materials

0.5 g/mL

Other packing materials

Use procedure below to determine density