Emptying the column

To remove packing material from the column:
1. Reverse the direction of flow through the column by connecting the

column outlet to the pump.

2. Remove the inlet connector by unscrewing the fine adjustment knob

from the inlet cap.

3. Screw the packing funnel into the inlet cap. Position a beaker under the

mouth of the funnel.

4. Start flow. Gradually increase the flow until the packed bed begins

to move. Continue flow until all packing material washes out of the
column.

5. Remove the funnel and clean the column. Make sure the threads are

clean.

6. Repack the column (see “Slurry-packing the column”, or “Dry-packing

the column”, for repacking instructions).

d. Installing the Column

The AP Column can be installed in:

Waters 650 Advanced Protein Purification System

Waters 625 system

Pharmacia

®

FPLC

®

systems

Other high performance liquid chromatography (HPLC) systems (refer
to Table 4 for adapters and unions used to connect the AP column to
selected systems)

Refer to the appropriate system user’s guide for specific column instal-
lation information.

Tightening fittings

Finger-tighten plastic fittings. Do not overtighten. Overtightening
may damage the threaded parts of the column. Overtightening does
not improve performance.

End plugs

Save the column end plugs for future column storage.

e. Equilibrating the Column

Before performing a test run, equilibrate the column:
1. Gradually increase the flow in 0.1 mL/min increments until you attain

the desired flow rate.

2. While equilibrating the column, check for voids; if present, gradually

reduce flow to 0 mL/min, Adjust the inlet connector until the void is
eliminated. (If this is unsuccessful, see “Eliminating Voids”).

3. Equilibrate the column with 10 to 20 column volumes of mobile phase,

or as recommended by the manufacturer of the packing material.

f. Eliminating Voids

Voids may develop in the column during normal use. To eliminate
voids:

1. Stop flow.

2. Disconnect the inlet column tubing from the system to allow displace-

ment of fluid while adjusting the plunger.

3. Deactivate the O-ring seal by turning the seal adjustment nut

counterclockwise.

4. Move the slide to the fine (down) position and turn it slowly to eliminate

the void.

If the fine adjustment knob is all the way down, back if off a few turns.
Move the slide to the coarse adjustment position and move the plunger
down a short distance. To eliminate the void, return the slide to the fine
adjustment position and turn it slowly.