Push air, Air cal, Hold air – RKI Instruments Gas Tracer User Manual
Page 87

Gas Tracer Operator’s Manual
Calibration Mode • 79
3. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and
press and release the POWER ENTER button. The unit will display
the current fresh air reading and prompt you to push the AIR▲
button.
NOTE:
If you have just performed a DRY fresh air adjustment, allow
the Gas Tracer to sample fresh air for 2 minutes before
continuing.
4. Be sure the Gas Tracer is in a fresh air environment.
5. Hold down the AIR▲ button. The unit will prompt you to continue
holding the AIR▲ button.
6. When the unit prompts you to release the AIR▲ button, release it.
The instrument will perform a fresh air adjustment on the ppm
sensor and return you to the AIR CAL menu item.
Performing a Dry Fresh Air Adjustment
This section describes how to perform a fresh air adjustment on the
PPM combustible gas sensor using the DRY setting. Use the DRY
setting with a zero air calibration cylinder for a biannual calibration.
NOTE:
It is recommended that you perform a fresh air adjust on the
ppm sensor using a zero air cylinder and the DRY setting
followed by a fresh air adjustment using the WET setting once
every 6 months to maintain the accuracy of the ppm
combustible gas reading. See “Performing a Wet Fresh Air
Adjustment” on page 78 for instructions to perform a WET
fresh air adjustment.
50
ppm
MAINTENANCE
CH
4
PUSH AIR
AIr
CAL
MAINTENANCE
HOLD AIR