30 • Measuring Mode, Normal Operation
Gas Tracer Operator’s Manual
NOTE:
Response time increases with the length of the sample hose.
Long sample hoses will require more time to show a response
at the Gas Tracer. The maximum sample hose length
recommended for the Gas Tracer is 50 feet. Consult RKI
Instruments, Inc. for longer sample hose lengths.
4. If a gas alarm occurs, take appropriate action. See “Responding to
Alarms” on page 41.
Using Optional Probes and Fittings
This section describes how to use the 8 meter hose with floating head and the
purge tee fitting.
Using the 8 Meter Hose with Floating Head
This probe is designed to be used in a tank or a well that may have water or
some other liquid at the bottom. To use the 8 meter hose with floating probe:
1. Start up the Gas Tracer as described in “Start Up, Normal Mode”
on page 20.
2. Connect the 8 meter hose with floating probe to the Gas Tracer’s
quick connect inlet fitting.
3. Slowly lower the probe into the area you wish to monitor. If the
probe hits water, it will float and continue to monitor the area.
Figure 15: 8 Meter Hose with Floating Head
Using the Purge Tee Fitting
The purge fitting is used to detect gas levels in a pressurized gas stream from
a pipeline or vessel being purged.
1. Start up the Gas Tracer as described in “Start Up, Normal Mode”