Bump dsp, Turning the bump test function on or off, Turning the pump off screen on or off – RKI Instruments Gas Tracer User Manual
Page 165

Gas Tracer Operator’s Manual
Using Maintenance Mode • 157
Turning the Bump Test Function On or Off
The BUMP DSP menu item is used to turn the bump test function on
or off. If set to OFF, the BUMP menu item in Calibration Mode and
Maintenance Mode does not appear. In addition, the BUMP--SET, BP
INTVL, BP RMNDR,
and BP EXPRD menu items will not display in
Maintenance Mode. The factory setting is OFF.
1. Use the AIR▲ or (SHIFT)▼ buttons to display the BUMP DSP
menu item.
2. Press and release the POWER ENTER button. The screen will
indicate the current setting.
3. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Turning the Pump Off Screen On or Off
The POFF DSP menu item is used to turn the Pump Off Screen on or
off. When set to ON, the Pump Off Screen appears in Display Mode
and the user can turn off the pump in order to conserve battery power.
If set to OFF (factory setting), the Pump Off Screen does not appear in
Display Mode and the user cannot turn the pump off.
WARNING: The Gas Tracer is not a gas monitoring device if the
pump is turned off.
MAINTENANCE
BUMP DSP
On
MAINTENANCE
BUMP DSP