Id disp, Bck lght, Turning the id display function on or off – RKI Instruments Gas Tracer User Manual
Page 175: Updating the lcd back light time setting

Gas Tracer Operator’s Manual
Using Maintenance Mode • 167
Turning the ID Display Function On or Off
With ID DISP set to ON, the User ID Screen and Station ID Screens
display during start up and in Display Mode. The ID’s can be updated
in Display Mode.
With ID DISP set to OFF (factory setting), the User ID Screen and
Station ID Screens do not display during start up or in Display Mode.
1. Use the AIR▲ or (SHIFT)▼ buttons to display the ID DISP menu
item. Press and release the POWER ENTER button. The screen will
indicate the current setting.
2. Use the AIR▲ or (SHIFT) ▼ button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Updating the LCD Back Light Time Setting
The BCK LGHT setting indicates the length of time the LCD
illuminates when you press any button. The minimum setting is 0
seconds; the maximum setting is 255 seconds. The factory setting is 30
seconds.
NOTE:
The backlight stays on during the entire warmup sequence.
1. Use the AIR▲ or (SHIFT)▼ buttons to display the BCK LGHT
menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.
2. Use the AIR▲ and (SHIFT)▼ buttons to display the desired setting.
On
MAINTENANCE
ID DISP
30
MAINTENANCE
BCK LGHT