Gas Tracer Operator’s Manual

Bump Testing • 209

lower right of the screen. The calibration time counted down
during a calibration initiated because of a failed bump test is
the difference between the bump test gas time and the
calibration time defined in the BUMP--SET menu item in
Maintenance Mode.

At the end of the calibration, the instrument displays the
results from both the bump test and the calibration. Use the
AIR▲ or (SHIFT)▼ button to scroll between the calibration/
bump test results, the bump test gas readings, and the
calibration gas readings.