beautypg.com

Bump 30, Bump – RKI Instruments Gas Tracer User Manual

Page 152

background image

144 • Using Maintenance Mode

Gas Tracer Operator’s Manual

6. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and

press and release the POWER ENTER button. The bump test gas
value will be displayed.

7. Before proceeding, confirm that the bump test gas value is the

same as the concentration in the calibration cylinder. If it is not,
adjust the bump test gas value by entering the AUTO CAL menu
item in the PPM CAL menu, changing the value there, and
reentering the BUMP menu.

8. Connect the humidifier from the demand flow regulator to the

rigid tube on the probe and quickly press and release the POWER
ENTER button. The unit will begin counting down from 30.

9. When AUTO CAL in the BUMP--SET menu item in Maintenance

Mode is set to OFF:

The instrument will display if the ppm combustible sensor
passed or failed the bump test with a P (pass) or an F (fail) to
the right of the gas.

1000


ppm

MAINTENANCE

CH

4

BUMP 30

P


ppm

MAINTENANCE

CH

4

BUMP