Alarm--p, Warning, Alarm – RKI Instruments Gas Tracer User Manual
Page 161
Gas Tracer Operator’s Manual
Using Maintenance Mode • 153
1. Use the AIR▲ or (SHIFT)▼ buttons to display the ALARM--P
menu item. Press and release the POWER ENTER button.
2. You may then select a channel. Use the AIR▲ or (SHIFT)▼ buttons
to select a channel. The combustible gas channel is shown for
example.
3. Press the POWER ENTER button to display the warning setting for
the combustible channel. It will be flashing.
4. Use the AIR▲ or (SHIFT)▼ buttons to change the warning setting.
5. Press POWER ENTER to save the setting. The alarm setting for the
combustible gas channel will then be displayed.
6. Use the AIR▲ or (SHIFT)▼ buttons to change the alarm setting.
- - -
%LEL
MAINTENANCE
CH
4
ALARM--P
10
%LEL
MAINTENANCE
CH
4
WARNING
50
%LEL
MAINTENANCE
CH
4
ALARM
- Eagle Series (89 pages)
- Eagle Series (66 pages)
- 01 Series Manual (33 pages)
- 01 Series Quik Start (2 pages)
- OX-94 (13 pages)
- FP-3 (16 pages)
- GasWatch 1 (18 pages)
- GasWatch 2 Manual (37 pages)
- GasWatch 2 Quik Start (2 pages)
- GP-01 (34 pages)
- GP-01 with self resetting alarms (33 pages)
- GP-204 (7 pages)
- NP-204 (13 pages)
- RI-215A (24 pages)
- SC-01 (45 pages)
- SC-01 (68 pages)
- Data Cal 2000 (88 pages)
- EAGLE 2 Manual (309 pages)
- Eagle 2 Quik Start (2 pages)
- GX-94 (52 pages)
- GX-94 (3 pages)
- SDM-E2 (89 pages)
- SDM-E2 (224 pages)
- Eagle 1 Manual (89 pages)
- Eagle 1 Quik Start (2 pages)
- Inert Gas Testing Eagle with Two Pumps (66 pages)
- Gas Tracer 10,000 ppm (120 pages)
- GX-2001 Manual (45 pages)
- GX-2001 Quik Start (2 pages)
- GX-2001 Canadian Version (43 pages)
- GX-2003 Manual (94 pages)
- GX-2003 Quik Start (2 pages)
- GX-2009 Manual (73 pages)
- GX-2009 Quik Start (2 pages)
- GX-2009 Supplement (2 pages)
- GX-2012 Manual (185 pages)
- GX-82 (22 pages)
- GX-82A (43 pages)
- GX-86 (22 pages)
- GX-86A (39 pages)
- GX-8000(PT0-098) (52 pages)
- GX-8000(H4-0050) (23 pages)
- RX-516 (36 pages)
- GX-2009 (2 pages)