204 • Bump Testing
Gas Tracer Operator’s Manual
11.After the last channel is calibrated, disconnect the calibration
tubing from the probe.
12.Use the AIR▲ and (SHIFT)▼ buttons to scroll to ESCAPE in the
ONE CAL menu.
13.Press and release the POWER ENTER button to exit ONE CAL.
14.With the ONE CAL menu option displayed, press the (SHIFT)▼
button until the NORMAL
menu option is displayed.
15.Press and release the POWER ENTER button to return to
Measuring Mode.
Bump Testing
NOTE:
The BUMP DSP menu item in Maintenance Mode is factory
set to OFF. The BUMP menu item will not appear unless
BUMP DSP
is set to ON. See “Turning the Bump Test
Function On or Off” on page 157 for instructions.
NOTE:
The Gas Tracer uses parameters defined in the BUMP--SET
menu item of Maintenance Mode while performing a bump
test. To view or update these parameters, see “Updating the
Bump Test Parameters” on page 158.
NOTE:
If the combustible gas channels are calibrated to a gas other
than methane, use an appropriate multigas cylinder.