22 • Start Up, Normal Mode
Gas Tracer Operator’s Manual
•
If the unit is due for calibration and CL EXPRD is set to NOT
USE, then the following screen displays.
The Gas Tracer cannot be used until a calibration has been
performed either by selecting AUTO CAL or ONE CAL in the
Maintenance Mode menu. See “Calibrating Using Auto
Calibration” on page 123 or “Calibrating Using Single
Calibration” on page 126 for calibration instructions.
•
If calibration is not due or CL EXPRD is set to NO EFFECT,
then the following screen appears for a few seconds indicating
when the next calibration is due.
6. If BP RMNDR is turned on (factory setting is OFF), the screen that
appears next depends on how BP EXPRD is set in the Maintenance
Mode Menu (page 162).
•
If the unit is due for a bump test and BP EXPRD is set to
CONFIRM (factory setting), then the following screen displays.
The alarm LED’s and buzzer will pulse several times. After
this, press the RESET SILENCE button until you hear a beep to
FAIL
C--LIMIT
2011
9
30
69d
NEXT CAL
tESt
B--LIMIT