One cal – RKI Instruments Gas Tracer User Manual
Page 77
Gas Tracer Operator’s Manual
Calibration Mode • 69
1. While in the calibration menu, use the AIR▲ or (SHIFT) ▼ button
to navigate to the ONE CAL menu option.
2. Press and release the POWER ENTER button to display the single
calibration menu. The %LEL sensor will be displayed first.
3. Use the AIR▲ or (SHIFT)▼ button to display the sensor you want
to calibrate (in this example the combustible gas % LEL sensor). If
you have a 5 sensor unit, the %vol sensor will be one of the choices.
It is shown below.
4. Press and release the POWER ENTER button. The single
calibration screen displays for the sensor you selected. The gas
reading flashes.
MAINTENANCE
ONE CAL
- - -
%LEL
MAINTENANCE
CH
4
ONE CAL
- - -
vol%
MAINTENANCE
CH
4
ONE CAL
0
%LEL
MAINTENANCE
CH
4
ONE CAL
- Eagle Series (89 pages)
- Eagle Series (66 pages)
- 01 Series Manual (33 pages)
- 01 Series Quik Start (2 pages)
- OX-94 (13 pages)
- FP-3 (16 pages)
- GasWatch 1 (18 pages)
- GasWatch 2 Manual (37 pages)
- GasWatch 2 Quik Start (2 pages)
- GP-01 (34 pages)
- GP-01 with self resetting alarms (33 pages)
- GP-204 (7 pages)
- NP-204 (13 pages)
- RI-215A (24 pages)
- SC-01 (45 pages)
- SC-01 (68 pages)
- Data Cal 2000 (88 pages)
- EAGLE 2 Manual (309 pages)
- Eagle 2 Quik Start (2 pages)
- GX-94 (3 pages)
- GX-94 (52 pages)
- SDM-E2 (89 pages)
- SDM-E2 (224 pages)
- Eagle 1 Manual (89 pages)
- Eagle 1 Quik Start (2 pages)
- Inert Gas Testing Eagle with Two Pumps (66 pages)
- Gas Tracer 10,000 ppm (120 pages)
- GX-2001 Manual (45 pages)
- GX-2001 Quik Start (2 pages)
- GX-2001 Canadian Version (43 pages)
- GX-2003 Manual (94 pages)
- GX-2003 Quik Start (2 pages)
- GX-2009 Manual (73 pages)
- GX-2009 Quik Start (2 pages)
- GX-2009 Supplement (2 pages)
- GX-2012 Manual (185 pages)
- GX-82 (22 pages)
- GX-82A (43 pages)
- GX-86 (22 pages)
- GX-86A (39 pages)
- GX-8000(PT0-098) (52 pages)
- GX-8000(H4-0050) (23 pages)
- RX-516 (36 pages)
- GX-2009 (2 pages)