beautypg.com

Normal, L.c. mode, Co disp off co disp on – RKI Instruments Gas Tracer User Manual

Page 110

background image

102 • Start Up, Leak Check Mode

Gas Tracer Operator’s Manual

Tracer’s quick connect inlet fitting.

NOTE:

If a probe is used when using the Gas Tracer in Leak Check
Mode, use the standard probe, not the bar hole probe.

3. Press and briefly hold down the POWER ENTER button. The

Normal Mode Select Screen displays.

4. Use the AIR▲ or (SHIFT)▼ button to scroll to the Leak Check

Mode Select Screen.

NOTE:

If the CO DISP parameter in Maintenance Mode is set to On,
the CO channel will appear in Leak Check Mode. If the CO
DISP

parameter is set to OFF (factory setting), the CO

channel will not appear in Leak Check Mode. All screens in
this section do not show the CO channel. If CO DISP is
turned on for your instrument, your screens may appear
different.

5. With the Leak Check Mode Select Screen displayed, press and

release the POWER ENTER button to begin the Leak Check Mode
startup sequence.

NOTE:

If no button is pressed for 20 seconds, the unit will proceed
into whichever mode is displayed.

0
20.9
0

vol%
%LEL

%

ppm

CH

4

O

2

NORMAL ?

CO

CO DISP OFF CO DISP On

0


5000

ppm

ppm

LEAK CK MODE

CH

4

L.C. MODE?

0

0
5000

ppm

ppm

ppm

LEAK CK MODE

CH

4

L.C. MODE?

CO