Normal, L.c. mode, Co disp off co disp on – RKI Instruments Gas Tracer User Manual
Page 110
102 • Start Up, Leak Check Mode
Gas Tracer Operator’s Manual
Tracer’s quick connect inlet fitting.
NOTE:
If a probe is used when using the Gas Tracer in Leak Check
Mode, use the standard probe, not the bar hole probe.
3. Press and briefly hold down the POWER ENTER button. The
Normal Mode Select Screen displays.
4. Use the AIR▲ or (SHIFT)▼ button to scroll to the Leak Check
Mode Select Screen.
NOTE:
If the CO DISP parameter in Maintenance Mode is set to On,
the CO channel will appear in Leak Check Mode. If the CO
DISP
parameter is set to OFF (factory setting), the CO
channel will not appear in Leak Check Mode. All screens in
this section do not show the CO channel. If CO DISP is
turned on for your instrument, your screens may appear
different.
5. With the Leak Check Mode Select Screen displayed, press and
release the POWER ENTER button to begin the Leak Check Mode
startup sequence.
NOTE:
If no button is pressed for 20 seconds, the unit will proceed
into whichever mode is displayed.