152 • Using Maintenance Mode
Gas Tracer Operator’s Manual
2. Press and release the POWER ENTER button. The screen will
indicate the current setting.
3. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Updating the Alarm Point Settings
The ALARM--P menu item allows you to update one or more
warning, alarm, STEL, and TWA (for CO) settings (the reading at
which the Gas Tracer recognizes these conditions).
For the combustible gas channel, it is not possible to adjust either the
Warning or Alarm settings to be higher than 60 %LEL. In addition, the
Warning setting cannot be adjusted higher than the Alarm setting for
all channels.
NOTE:
It is not necessary to adjust every setting for every channel
when you enter this menu item. All settings are shown here
for example.
Table 7: Factory Set Alarm Points
Channel
Warning
Alarm
STEL
TWA
Combustible Gas
10% LEL
50% LEL
N/A
N/A
Oxygen
19.5% decreasing
23.5% increasing
N/A
N/A
CO
25 ppm
50 ppm
200 ppm
25 ppm
300
MAINTENANCE
LG INTVL