Interface rs 232, A. query – Heidolph Synthesis 1 User Manual
Page 100

100
E
17. Interface RS 232
Interface commands
a. Query
SZn? with n=1...4
nominal temperature, zone n
answer: SZn=xxx
SD?
nominal
rotational
speed
answer: SD=xxxx
ST?
nominal time, timer elapse
answer: ST=xxx
BT?
running
down,
timer
answer: BT=xxx
AZn? with n=1...4
actual temperature, zone n
answer: AZn=xxx
AEn? with n=1...4
actual value, external sensor, zone n
answer: AEn=xxx
AD?
actual
rotational
speed
answer: AD=xxxx
SB?
operational
status
answer: SB=x
Answer
heater
speed
Ramp Active
Identic
Bit
b0
b1
b2
b3
SB=0
off off off off
SB=1
on off off off
SB=2
off on off off
SB=3 on
on
off
off
SB=4
off off on off
SB=5
on off on off
SB=6 off
on
on
off
SB=7
on on on off
SB=8
off off off on
SB=9
on off off on
SB=10
off on off on
SB=11 on
on
off
on
SB=12
off off on on
SB=13
on off on on
SB=14 off
on
on
on
SB=15
on on on on
FN?
error
number
answer: FN=x