Operating personnel’s obligations, Personnel qualification changes to the unit – Heidolph Automatic Module Distimatic User Manual
Page 7
Safety Instructions
12
13
Safety Instructions
Operating instructions
These operating instructions as well as the operating instructions for
the Hei-VAP or Hei-VAP Industrial must be followed, read and
understood by all persons working with the device (acknowledgment
of the safety instructions is particularly important).
Operating personnel’s obligations
Verify that the distillation material can be evaporated safely and that the distillation
residue is not explosive.
Ensure that work which uses naked flames is not carried out in the vicinity of the
device (risk of explosion).
Ensure that the flow rate when removing liquids with flammable materials remains at
< 1 m/s (electrostatic charge; risk of ignition).
Ensure that gases classified in explosion group IIC are not produced by materials or
during chemical reactions, e.g. hydrogen.
Ensure that no devices are operated or assembled which are emission or radiation
sources (electromagnetic waves) for the frequency range (3×1011 Hz to 3×1015 Hz).
Ensure that no devices are operated or assembled which constitute emission or
radiation sources for ionizing radiation or in the ultrasonic range.
Ensure that no adiabatic compression or shock waves occur (shock wave combustion).
Ensure that the use of materials which pose a potential risk of uncontrolled energy
release is prohibited by an associated pressure increase (exothermic reaction;
spontaneous combustion of dusts).
Ensure that glass surfaces are only wiped with damp cloths.
Ensure that ground joint connections are clean and sit securely attached.
Please wear the appropriate protective clothing when working on the machine
(protective glasses and if necessary, safety gloves).
Ensure that sufficient heat transfer media are used.
Ensure that the maximum excess pressure of the coolant equals no more than 1 bar
in the condenser.
Ensure that all tube connections are properly connected to ensure safe operation.
Ensure that all cables and tubes are correctly laid and are located outside the
operating and danger zone.
Please avoid putting pressure on the display.
Personnel qualification
Changes to the unit
The target group for this device is qualified personnel. The device may only be operated
by individuals who have been instructed in its proper use by qualified personnel.
No unauthorized changes may be made to the unit. No parts may be used which have not
been approved by the manufacturer. Unauthorized changes result in the CE declaration of
conformity losing its validity, and the appliance may no longer be operated.
The manufacturer is not liable for any damage, danger or injuries that result from
unauthorized changes or from operating the unit other than described in this manual.