Installation

Installation

54

55

Connect PTFE tube Ø 6 mm to

connection (1) and tighten until hand-

tight.

Route PTFE tubes and flask sensor into the evaporating flask and secure the multiple

screw connection at connection (4) until hand-tight.

Multiple screw connection

Route PTFE tube Ø 8 mm through connection (2) (only on Distimatic including

automatic drainage of residue) and PTFE tube Ø 6 mm through connection (3) and

secure in place so it is hand-tight.

The immersion depth of the PTFE tubes and the flask sensor in the evaporating flask

may be determined by loosening the respective screw cap slightly at the multiple

screw connection and by pulling or pushing the PTFE tube or the flask sensor.

Re-secure the screw caps on the multiple screw connection so they are hand-tight

again afterwards.

1

2

5

3

7

6

Optimization of the distillation process

Refill evaporating flask only up to one third. Position the flask sensor

accordingly.

Use large evaporating flasks (3-5 liter variants) whenever possible.
Use the following recommended immersion depths:

The end of the PTFE tube (5) Ø 6 mm varies. Flask sensor (6) ends just

above the required maximum fill level. PTFE tube (7) Ø 8 mm (only on

Distimatic including automatic drainage of residue) ends approximately

2 – 3 mm above the bottom of the evaporating flask.
This way your distillation process will become much faster.