Maintenance, Cleaning, Service

Maintenance, Cleaning, Service

WARNING

Hot surfaces!
Risk of scalding and burns.

Allow the glass equipment to cool down.
Allow media to cool down.

WARNING

Hazardous materials and solvents!
There are risks posed by the leakage of hazardous substances and

solvents from locations that are not properly sealed.

Check whether all screw caps have been tightened properly at the

threads to seal them off.

Route all PTFE tubes into the vessels intended for them and fasten

them securely to the vessel.

Do not expose operating personnel to harmful vapors or gases.

Extract waste air using a vapor outlet.

Ensure that the separator for the power supply is easily accessible at

all times.

Cleaning agents
To ensure thorough cleaning, replace contaminated cleaning agents

during the process for clean cleaning agents, if necessary.

Ensure that the device is in the Sensor-Mode and in the operating state process

stopped.

Allow residue pump and condensate pump to continue running for an additional

10 seconds until the residue and condensate circuits are drained.

Product storage circuit

Route the PTFE tube which leads to the product storage vessel into a vessel

containing cleaning agents (e.g. solvent).

Create a vacuum.
Now follow the chapters entitled Start-Up - Set parameters to convey cleaning

agent into the flask.

Remove the PTFE tube from the vessel containing cleaning agents and direct into the

atmosphere. Air will be conveyed into the evaporating flask.

Wait 10 seconds until the product storage circuit is emptied.

Condensate circuit

To create atmospheric pressure within the evaporator system, ventilate the rotary

evaporator.

Hei-VAP: Remove the vacuum attachment or vacuum tube (glassware set G1) from

the uppermost condenser opening; Hei-VAP Industrial: Remove vacuum fitting.

Then press the start/stop key. The message please preset flask-level manually

appears in the display. Disregard this message.

Use the uppermost condenser opening to fill in the cleaning agent. The cleaning agent

will be discharged by the condensate pump.

Wait 10 seconds until the cleaning agent has drained into the collector.
To trigger the level sensor of the collector manually, position your index finger on the

level sensor (between the collector and the metal piece onto which the level sensor is

fastened). The orange level sensor LED will light up and the collector will be drained.

Repeat this process until the collector is completely empty.
Wait 10 seconds until the condensate circuit is emptied.

Residue circuit

Pump cleaning agents, as outlined under „Product storage circuit” into the

evaporating flask.

Route the PTFE tube which leads to the residue vessel into an empty vessel.
Press the „drain residue” key. Cleaning agent will be drained by the residue pump.
Wait 10 seconds until the residue circuit is emptied.

Flask sensor

Remove the flask sensor out of the evaporating flask and glassware set and clean

with a cloth drenched with cleaning agent.

Disassembly

WARNING

Hot surfaces!
Risk of scalding and burns.

Allow the glass equipment to cool down.
Allow media to cool down.

Disassembly of the Hei-VAP and Hei-VAP Industrial

See also the operating instructions of Hei-VAP or Hei-VAP Industrial.