Operating states

Operating state

Description

process stopped

Automatic mode is disabled.
Basic condition after activation of device or end of

automatic mode.

Status display

Automatic mode is activated.

endprocess

Automatic mode in final process.
If the refill sensor does not detect any medium during

the refill, the final process is initiated. The remaining

media is distilled.
Following, the mains sockets for the rotary evaporator

and the vacuum pump are deactivated at the control

box. The vacuum circuit is ventilated. The condensate

pump pumps out the condensate.

Parameter overview

Set parameters – Sensor-Mode

All parameters of the following table are settable. The settable criteria of a parameter

(parameter settings) are marked in the following with the symbols „>…<”.

Select operational mode

After switching on the unit you can choose between two different operational modes.
Select Sensor-Mode by pressing the set key and the arrow key <.

Select Time-Mode by pressing the set key and the arrow key >.
Once you selected the Sensor-Mode a S will be visible on the display.

Once you selected the Time-Mode a T will be visilble on the display.
Changing modes is only possible after resetting the unit.

Parameter

Description

Start/stop

To gain access to the menu.

Refill to start

initial filling

a. The key refill starts the pre-programmed operation.
b. The key parameter allows for pre-setting all

parameters.

Time to refill >…<

Time between refilling of the medium. If the preset time

has expired, the flask sensor checks whether there is

still any medium in the evaporating flask and triggers

product refilling if needed.

Extra time until drain off

The residue in the evaporating flask is drained by the

residue pump if the evaporation process has stagnated

for the time period set in the parameter Extra time until

drain off.

Drain time

cond >…<

resi >…<

Cond = Drainage of condensate

Please enter the required time to drain condensate from

the collector

Resi = Drainage of residue

Please enter the required time to drain residue from the

evaporating flask.

ENDPROCESS
evaporate time >…<

Pre-program the time frame for the unit to power off all

periphery after the last cycle has been executed.

ENDPROCESS

draining
cond >…<

resi >…<

Prior to powering off, the collector (condensate) and the

evaporating flask (residue) can be drained. You have the

option to allow for a final drainage (ON) or not (OFF).

Residue pump during

process >…<

To drain the residue this option must be set ON.

Refill to start

initial filling

Start the process by pressing the refill key.

Overflow Collection

Vessel >…<

Activate the overflow power-off by pressing the ON key

or not (OFF).

Mo 09:00
2006-01-27

Date and time display format. Here you can set the

current date and time. Since error codes are always

stamped with time, we recommend you to keep the date

up to date in order to facilitate error analysis.

Drainage of condensate or residue can be performed manually by pressing the refill /

drain key (condensate) and drain residue key (residue).