Operation, Switching the control box on operating panel – Heidolph Automatic Module Distimatic User Manual
Page 40
![background image](/manuals/561954/40/background.png)
Start-Up
Start-Up
78
79
Switch on the control box using the mains switch (1) (the cut-out switch must not be
activated).
After a few seconds, the operating state process stopped appears in the display of the
operating panel. The device is ready for operation.
The device is operated via the operating panel. The operating elements are described
below.
Switching the control box on
Operating panel
Control elements
Automatic mode is not activated
Automatic mode is not active after switching the device on.
WARNING
Hot surfaces!
Risk of scalding and burns.
Allow the glass devices to cool down sufficiently.
Allow the media to cool down.
WARNING
Hazardous materials and solvents!
There are risks posed by the leakage of hazardous substances and
solvents from locations that are not properly sealed.
Check whether all screw caps have been tightened properly at the
threads to seal them off.
Guide all PTFE tubes into the vessels intended for them and fasten
them securely to the vessel.
Do not expose operating personnel to harmful vapors or gases.
Extract waste air using a vapor outlet.
Ensure that the separator for the power supply is easily accessible at
all times.
Operation
Control element
Description
Display
Displays the operating states,
parameters and messages.
Arrow keys
up/down
a. Select individual parameters (within the menu).
b. Change the characters/digits of a parameter setting
(within a parameter).
Arrow keys
right/left
a. Navigate to the next character/digit of a parameter
setting (within a parameter).
b. Navigate between parameter settings (within a
parameter with several parameter settings).
set key
Select the first parameter setting (within a parameter
with several parameter settings).
set and apply keys
(in combination)
a. Select individual parameter settings (press „set”
for at least 2 seconds) and select with the cursor
(apply) (within a parameter).
b. Save changed parameter setting (press apply
for at least 2 seconds and then set for at least 2
seconds; within a parameter).
parameter key
a. Open menu.
b. Leave menu.
drain residue key
Residues, if any, in the evaporating flask, are removed
manually. The residue pump drains the residue as
long as you keep the key pressed (only on Distimatic
including automatic drainage of residue and if
automatic mode is not activated).
refill/drain key
Collector is manually emptied
(if automatic mode is not activated).
start/stop key
Automatic mode is enabled or disabled.
start: If an operating pump is connected, vacuum will
be applied to the system.
stop: The system will be ventilated.