beautypg.com

Multiple switched socket and control box – Heidolph Automatic Module Distimatic User Manual

Page 30

background image

Installation

Installation

58

59

Mount 3/2 way valve on stand rod (14) and secure using screws (15).

Connect vacuum tube to connection (18).

Connect vacuum tube to connections (16) and (17).

Vacuum circuit

in direction of vacuum pump

14

17

16

15

DANGER

Danger to life due to dangerous electrical voltage!
Risk of serious injury.
Electrical residual energy remains in lines, equipment and devices after

shutting down the device.

Point out the power sockets to the operating

personnel. Disconnect the power plugs from the energy

source in order to establish zero potential.

Only allow qualified electricians to perform

work on the electrical supply system.

Disconnect the device from the mains supply. Regularly check

the electrical equipment (mains cables) of the device.

Replace damaged power cords. Regularly check all power

cords for damages during repair and maintenance work.

Only set up or disconnect connections when discon-

nected from the power supply.

Multiple switched socket and control box

18

Connection principle

See chapter entitled „Function and connection principle” - Hei-VAP

Distimatic”- „Connection principle of power supply”.