Maintenance cleaning, Housing circuits and flask sensor – Heidolph Automatic Module Distimatic User Manual
Page 46
91
Maintenance, Cleaning, Service
Error Codes
90
Error code
Cause
Consequence
Measure
Error 4:
Unable to
drain residue
Automatic drainage
of residue active:
The residue in the
evaporating flask can
no longer be drained.
a. Suction and pressure
height of the residue
pump are exceeded.
The flask sensor
triggers safe
shutdown and
initiates the
final process (in
the event of an
error). The error
code appears.
a. Compare suction
and pressure height
of the residue pump
with the data of the
maximumsuction
and pressure height
given in chapter
„Technical data” and
adjust the heights
correspondingly.
b. Flow rate of the
residue pump is
set too low.
b. Adjust the flow rate
of the residue pump
using the rotary
knob to 100 %.
c. The residue
pump is dirty.
c. Clean the residue
pump (see chapter
entitled „Cleaning”
- „Circuit and
flask sensor” -
„Residue circuit”).
d. The viscosity of the
medium is too high.
d. Adjust the viscosity
of the medium for
the new evapo-
ration process.
e. Cable break in the
flask sensor.
e. Contact Technical
Service.
f. The residue pump is
defective.
f. Contact Technical
Service.
g. PTFE tube Ø 8 mm
for the automatic
drainage of residue
(only on Disti-
matic including
automatic drainage
of residue)incor-
rectly positioned.
g. Correct the position
of the PTFE tube
Ø 8 mm for the
automatic drainage
of residue (see
chapter „Instal-
lation”).
Maintenance
Cleaning
The device is maintenance-free.
Is repair required?
Contact Heidolph Instruments. Repair must be performed by a specialist authorized
by Heidolph Instruments.
Housing
Circuits and flask sensor
CAUTION
Damage to property as a result of corrosive cleaning agents!
Damages to the surface of the device.
To clean the housing and device surface, use a damp cloth (mild
soap solution). Do not use cotton waste, scouring agents, acids,
alkaline solutions, chlorine bleach, chlorine-based cleaning products
or products which contain metallic components.
Servicing the Hei-VAP and Hei-VAP Industrial
See also the operating instructions of Hei-VAP or Hei-VAP Industrial.
DANGER
Risk posed by electrical voltage!
Risk of damage to health and to the device.
Remove mains plug before opening the housing.
WARNING
Glass breakage!
Serious injuries due to shattering glass and glass breakage.
Check glass devices for damage (breakage, cracks etc.).
Only use glass equipment in perfect condition.
Monitor pressure conditions in the glass attachment.
Work carefully.