Maintenance, Cleaning, Service

Error Codes

90

Error code

Cause

Consequence

Measure

Error 4:

Unable to

drain residue

Automatic drainage

of residue active:

The residue in the

evaporating flask can

no longer be drained.
a. Suction and pressure

height of the residue

pump are exceeded.

The flask sensor

triggers safe

shutdown and

initiates the

final process (in

the event of an

error). The error

code appears.

a. Compare suction

and pressure height

of the residue pump

with the data of the

maximumsuction

and pressure height

given in chapter

„Technical data” and

adjust the heights

correspondingly.

b. Flow rate of the

residue pump is

set too low.

b. Adjust the flow rate

of the residue pump

using the rotary

knob to 100 %.

c. The residue

pump is dirty.

c. Clean the residue

pump (see chapter

entitled „Cleaning”

- „Circuit and

flask sensor” -

„Residue circuit”).

d. The viscosity of the

medium is too high.

d. Adjust the viscosity

of the medium for

the new evapo-

ration process.

e. Cable break in the

flask sensor.

e. Contact Technical

Service.

f. The residue pump is
defective.

f. Contact Technical

Service.

g. PTFE tube Ø 8 mm

for the automatic

drainage of residue

(only on Disti-

matic including

automatic drainage

of residue)incor-

rectly positioned.

g. Correct the position

of the PTFE tube

Ø 8 mm for the

automatic drainage

of residue (see

chapter „Instal-

lation”).

Maintenance

Cleaning

The device is maintenance-free.
Is repair required?

Contact Heidolph Instruments. Repair must be performed by a specialist authorized

by Heidolph Instruments.

Housing

Circuits and flask sensor

CAUTION

Damage to property as a result of corrosive cleaning agents!
Damages to the surface of the device.

To clean the housing and device surface, use a damp cloth (mild

soap solution). Do not use cotton waste, scouring agents, acids,

alkaline solutions, chlorine bleach, chlorine-based cleaning products

or products which contain metallic components.

Servicing the Hei-VAP and Hei-VAP Industrial

See also the operating instructions of Hei-VAP or Hei-VAP Industrial.

DANGER

Risk posed by electrical voltage!
Risk of damage to health and to the device.

Remove mains plug before opening the housing.

WARNING

Glass breakage!
Serious injuries due to shattering glass and glass breakage.

Check glass devices for damage (breakage, cracks etc.).
Only use glass equipment in perfect condition.
Monitor pressure conditions in the glass attachment.
Work carefully.