Maintenance
InjectMan

®

4

English (EN)

82

8

Maintenance

8.1

Replacing fuses

The fuse holder is located between the mains connection socket and the mains power
switch. The fuses may only be replaced with the same type of fuse.

1. Disconnect the mains plug.

2. Pull the fuse holder out completely.

3. Replace the defective fuse.

4. Insert the fuse holder.

8.2

Cleaning

Prerequisites

• Clean at least every 4 weeks.

1. Wipe the painted parts and the aluminum surfaces with a cloth and mild detergent.

2. Wipe with a dry cloth afterwards.

DANGER! Electric shock.

Switch off the device and disconnect the power plug before starting
maintenance or cleaning work.

DANGER! Electric shock as a result of penetration of liquid.

Switch off the device and disconnect the power plug before starting cleaning
or disinfection work.

Do not allow any liquids to penetrate the inside of the housing.

Do not spray clean/spray disinfect the housing.

Only plug the device back in if it is completely dry, both inside and outside.

NOTICE! Damage from the use of aggressive chemicals.

Do not use any aggressive chemicals on the device or its accessories, such as
strong and weak bases, strong acids, acetone, formaldehyde, halogenated
hydrocarbons or phenol.

If the device has been contaminated by aggressive chemicals, immediately
clean it by means of a mild cleaning agent.