4 joystick, 5 software and parameters, 2 error messages – Eppendorf InjectMan 4 User Manual
Page 80: 1 warnings
Troubleshooting
InjectMan
®
4
English (EN)
80
7.1.4
Joystick
7.1.5
Software and parameters
7.2
Error messages
7.2.1
Warnings
The display does not
show anything or the
device cannot be
activated although the
device is connected.
• The power cable or power
plug is loose.
• The device is switched off.
Check the power cable or
the power plug.
Switch the device on.
Symptom/message
Cause
Remedy
Outer movement range
of the joystick does not
work.
• Outer movement range is
deactivated.
In the
Installation
menu, set
the value for
Dyn-mode
to
ON
.
Symptom/message
Cause
Remedy
Parameters are not
accessible for certain
operations.
–
Enter parameters again.
Carry out a reset and reset
all parameters to the works
settings.
Readjust device.
Symptom/message
Cause
Remedy
WARNING 1
• FemtoJet is not connected.
Disconnect the data cable to
the FemtoJet and connect it
again.
Switch on the FemtoJet.
WARNING 3
• X-module is not connected.
Switch off the device.
Connect the plug of the
X-module to the control
board and tighten.
Switch the device on.
WARNING 4
• Y-module is not connected.
Switch off the device.
Connect the plug of the
Y-module to the control
board and tighten.
Switch the device on.
Symptom/message
Cause
Remedy