6.14.2

Setting the parameter for

Clean

1. Set

Home speed

.

2. Set

Clean distance

.

3. Close the menu.

6.14.3

Execute the

Clean

function

Prerequisites
• Parameters are set for

Clean

.

Clean

is assigned to a softkey.

• The capillary is moved out of the medium with the set parameters.

6.14.4

Terminate the

Clean

function.

• The function is terminated.
• The application screen appears.

Parameter

Value

Range of values Increment Standard

Home speed

Set the speed of the clean
function

5 – 10000

5

7500

Clean distance

Set distance in μm

0 – 20000

5

10000

Set the

Clean distance

above the medium boundary