11 inject function, 1 inject menu and parameters – Eppendorf InjectMan 4 User Manual
Page 63
Operation
InjectMan
®
4
English (EN)
63
6.11
Inject
function
This function performs an automatic injection movement. You can set the injection speed,
the injection time, the injection movement and fix the distance between the capillary and
the lower limit.
6.11.1
Inject
menu and parameters
Abb. 6-2:Menu – Inject
Fig. 6-2:
Menu –
Inject
Parameter
Value
Range of values Increment Standard
Injection speed
Set the injection speed in μm
per second
5 – 10000
5
300
Synchr. inject.
Set the synchronized time of
injection
MOVE
IMMEDIATE
LIMIT
PRESSURE
–
LIMIT
Search+limit
Change the lower limit and
the position of the capillary.
The distance between the
capillary and the limit
remains constant
OFF/ON
–
OFF
Injection axial
Switch the axial movement of
the function on or off
OFF/ON
–
On
Parameter
Range of values Function
Synchr. inject.
MOVE
Injection movement without injection pressure (when
using non-connected injectors)
IMMEDIATE
Injection pressure starts with the injection movement
LIMIT
Injection pressure at the end of the injection distance
(
Z-axis Limit
)
PRESSURE
Injection pressure without injection movement