6.10

Speed

function

In the

Speed

menu you can set the speed of the working range and the speed at which

certain positions are approached or movements executed.

6.10.1

Speed

menu and parameters

Abb. 6-1:Menu – Speed

Fig. 6-1:

Menu –

Speed

6.10.2

Setting the parameters for

Speed

1. Select the desired parameter.

2. Set the value for the parameter.

3. Close the menu.

Parameter

Value

Range of values Increment Standard

Coarse

Set the speed in μm per
second

5 – 10000

5

7500

Fine

Set the speed in μm per
second

5 – 1000

5

1000

X-fine

Set the speed in μm per
second

0 – 100

1

100

Injection speed

Set the speed in μm per
second

5 – 10000

5

300

Step inj. speed

Set the speed in μm per
second

5 – 10000

5

300

Position speed

Set the speed in μm per
second

5 – 10000

5

1500

Home speed

Set the speed in μm per
second

5 – 10000

5

7500

If a value of 0 is set for the

X-fine

parameter, then the change between the

Fine

and

X-fine

working ranges is deactivated.