10 speed function, 1 speed menu and parameters, 2 setting the parameters for speed – Eppendorf InjectMan 4 User Manual
Page 62
Operation
InjectMan
®
4
English (EN)
62
6.10
Speed
function
In the
Speed
menu you can set the speed of the working range and the speed at which
certain positions are approached or movements executed.
6.10.1
Speed
menu and parameters
Abb. 6-1:Menu – Speed
Fig. 6-1:
Menu –
Speed
6.10.2
Setting the parameters for
Speed
1. Select the desired parameter.
2. Set the value for the parameter.
3. Close the menu.
Parameter
Value
Range of values Increment Standard
Coarse
Set the speed in μm per
second
5 – 10000
5
7500
Fine
Set the speed in μm per
second
5 – 1000
5
1000
X-fine
Set the speed in μm per
second
0 – 100
1
100
Injection speed
Set the speed in μm per
second
5 – 10000
5
300
Step inj. speed
Set the speed in μm per
second
5 – 10000
5
300
Position speed
Set the speed in μm per
second
5 – 10000
5
1500
Home speed
Set the speed in μm per
second
5 – 10000
5
7500
If a value of 0 is set for the
X-fine
parameter, then the change between the
Fine
and
X-fine
working ranges is deactivated.