6.9
Horizontal limit
For the X-axis, a limit can be defined for a horizontal injection. This prevents the capillary
from moving through the sample.
6.9.1
Defining the horizontal limit
1. In the
Installation
menu, select the
Angle
submenu.
2. Set the injection angle to 0° and save with
Enter
.
3. Close the menu.
The application screen now displays
X-axis Limit
.
4. Move the capillary to the desired final position for the X-axis.
5. Save the lateral limit with
X-axis Limit
.
• The X-coordinate is marked with
L
.
• The value for the limit (
Limit X
) is displayed.
•
X-axis Limit
is selected.
• The capillary cannot be moved any further to the side.
6.9.2
Deleting the horizontal limit
1. Press
X-axis Limit
.
The limit is deactivated.
2. In the
Installation
menu, select the
Angle
submenu.
3. Reset the injection angle to the operating angle and save with
Enter
.
• The lateral limit is deleted.
• The application screen displays
Back
again.