61

6.9

Horizontal limit

For the X-axis, a limit can be defined for a horizontal injection. This prevents the capillary
from moving through the sample.

6.9.1

Defining the horizontal limit

1. In the

Installation

menu, select the

Angle

submenu.

2. Set the injection angle to 0° and save with

Enter

.

3. Close the menu.

The application screen now displays

X-axis Limit

.

4. Move the capillary to the desired final position for the X-axis.

5. Save the lateral limit with

X-axis Limit

.

• The X-coordinate is marked with

L

.

• The value for the limit (

Limit X

) is displayed.

X-axis Limit

is selected.

• The capillary cannot be moved any further to the side.

6.9.2

Deleting the horizontal limit

1. Press

X-axis Limit

.

The limit is deactivated.

2. In the

Installation

menu, select the

Angle

submenu.

3. Reset the injection angle to the operating angle and save with

Enter

.

• The lateral limit is deleted.

• The application screen displays

Back

again.