6.7.5

Overwriting a stored position

Prerequisites

• A position is stored.

1. Press the softkey for a different position.

• You will move to the position.

2. When the position has been reached, press and hold the softkey for the position to be

overwritten.

• The old position is overwritten with the current coordinates.

3. Press the softkey.

• An acoustic signal will sound.

• In the softkey status field a filled circle is displayed.

• The stored position is displayed in the coordinate field.

6.8

Vertical limits

For the Z-axis, a lower and upper limit can be defined. This prevents the capillary coming
into contact with the bottom of the Petri dish or moving against the condenser of the
microscope adapter.

• Lower limit –

Z-axis Limit

• Upper limit –

Upper limit

6.8.1

Defining the lower limit

Prerequisites

• An application has been selected.

1. Position the capillary a little way above the slide.

2. Press

Z-axis Limit

.

• The Z-coordinate is marked with

L

.

• The value for the limit (

Limit Z

) is displayed.

Z-axis Limit

is selected.

• The capillary cannot be moved any lower.