6.7.5
Overwriting a stored position
Prerequisites
• A position is stored.
1. Press the softkey for a different position.
• You will move to the position.
2. When the position has been reached, press and hold the softkey for the position to be
overwritten.
• The old position is overwritten with the current coordinates.
3. Press the softkey.
• An acoustic signal will sound.
• In the softkey status field a filled circle is displayed.
• The stored position is displayed in the coordinate field.
6.8
Vertical limits
For the Z-axis, a lower and upper limit can be defined. This prevents the capillary coming
into contact with the bottom of the Petri dish or moving against the condenser of the
microscope adapter.
• Lower limit –
Z-axis Limit
• Upper limit –
Upper limit
6.8.1
Defining the lower limit
Prerequisites
• An application has been selected.
1. Position the capillary a little way above the slide.
2. Press
Z-axis Limit
.
• The Z-coordinate is marked with
L
.
• The value for the limit (
Limit Z
) is displayed.
•
Z-axis Limit
is selected.
• The capillary cannot be moved any lower.