6.15
PiezoXpert
function
6.15.1
PiezoXpert
menu and parameters
Abb. 6-6:Menu – PiezoXpert
Fig. 6-6:
Menu –
PiezoXpert
Parameter
Value
Range of values Increment Standard
Synchr.
PiezoXp.
Switch synchronization
between InjectMan 4 and
PiezoXpert on or off
OFF/ON
PiezoXp. speed
Set the advance rate of the
manipulator in μm/s while
the PiezoXpert triggers a
pulse.
5 – 10000
5
300
PiezoXp. dist.
Set the length of the
movement step of the
manipulator in μm while the
PiezoXpert triggers a pulse.
1 – 2000
1
20
PiezoXp. axial
Axial movement on or off
(corresponds to the set
injection angle)
On = movement in the X and
Y-axis.
Off = movement in the X-axis
ON/OFF
–
OFF
See also other documents in the category Eppendorf Equipment:
- epMotion 96 (76 pages)
- epMotion 5070 (100 pages)
- epMotion 5075 (130 pages)
- Centrifuge 5427 R (64 pages)
- Centrifuge 5427 R (104 pages)
- White Paper 14 (8 pages)
- Rolling Cabinet (34 pages)
- Mastercycler nexus (118 pages)
- Mastercycler nexus (142 pages)
- Concentrator plus (New Design) (48 pages)
- Concentrator plus (43 pages)
- Easypet 3 (38 pages)
- Xplorer (74 pages)
- Xplorer Adjustment (26 pages)
- AF2200 Plate Reader (72 pages)
- AF2200 Plate Reader (78 pages)
- G0.5 µPlate (32 pages)
- BioSpectrometer basic (104 pages)
- BioSpectrometer kinetic (106 pages)
- BioSpectrometer fluorescence (102 pages)
- Micro Test Tubes (5 pages)
- Microplates (10 pages)
- PiezoXpert (34 pages)
- Eporator (38 pages)
- MiniSpin (20 pages)
- MiniSpin (25 pages)
- Centrifuge 5702 (32 pages)
- 5702 Centrifuge (27 pages)
- 5702 Centrifuge (32 pages)
- C5702 RH Centrifuge (32 pages)
- 5418 Centrifuge (80 pages)
- 5418 Centrifuge (48 pages)
- 5424 Centrifuge (71 pages)
- 5424 Centrifuge (44 pages)
- 5430 Centrifuge (88 pages)
- 5430 Centrifuge (130 pages)
- 5804 Centrifuge (95 pages)
- 5804 Centrifuge (127 pages)
- 5804 Centrifuge (129 pages)
- TransferMan4 r (102 pages)
- TransferMan4 m (96 pages)
- InjectMan NI 2 (60 pages)
- InjectMan NI 2 (16 pages)
- PatchMan NP 2 (53 pages)