12 step injection function, 1 step injection menu and parameters – Eppendorf InjectMan 4 User Manual
Page 65

Operation
InjectMan
®
4
English (EN)
65
6.12
Step injection
function
With this function you can execute a straight injection over a defined distance.
Step
injection
can only be executed when a FemtoJet is connected.
6.12.1
Step injection
menu and parameters
Abb. 6-3:Menu – Step injection
Fig. 6-3:
Menu –
Step injection
Parameter
Value
Range of values Increment Standard
Step injection
Switch the function on or off
OFF/ON
–
OFF
Step inj. speed
Set the injection speed in μm
per second
5 – 10000
5
300
Step inj. dist.
Set the injection path in μm
1 – 2000
1
20
Step inj. axial
Switch the axial movement of
the capillary on or off
OFF/ON
–
On
Step inj. back
Switch the backward
movement of the capillary on
or off
OFF/ON
–
On
Synchr. inject.
Set the time of injection
MOVE
IMMEDIATE
LIMIT
PRESSURE
–
LIMIT
Parameter
Range of values Function
Synchr. inject.
MOVE
Injection movement without injection pressure
IMMEDIATE
Injection pressure starts with the injection movement
LIMIT
Injection pressure at the end of the injection distance
(
Step inj. dist.
)
PRESSURE
Injection pressure without injection movement