Installation

InjectMan

®

4

English (EN)

29

4.6

Inserting the capillary into the capillary holder

1. Loosen the knurled screw on the angle

head.

2. Insert the capillary holder into the

clamp.

3. Align the capillary holder in such a way

that the capillary tip is located approx.
20 mm above and approx. 20 mm from
the outside of the operating point.

4. Place the positioning aid on the capillary

holder and tighten.

5. Tighten the knurled screw.

WARNING! Risk of injury due to flying capillaries and glass splinters.
If exposed to high pressures, capillaries may detach themselves from the grip
heads and become projectiles.
Capillaries can crack as a result of incorrect handling.

Wear protective goggles.

Never aim capillaries at people.

Use capillaries with an outer diameter that matches the grip head
specifications.

Always mount / dismount capillaries when they are depressurized.

Mount the capillary correctly in the grip head.

Do not touch the capillary with the Petri dish or other objects.

NOTICE! Mechanical damage to the motor modules.
Excessive load leads to increment errors or destruction of the drive.

Do not drive the modules against mechanical obstructions.

Do not hold any objects near the modules.

Load the motor module with a maximum of 100 g.

Standard capillary: Only use the 0 grip head with capillaries with an outer
diameter of 1.0 mm to 1.1 mm. If you would like to use other capillaries, order
the matching grip head.