7 troubleshooting, 1 general errors, 1 motor module – Eppendorf InjectMan 4 User Manual
Page 79: 2 capillary, 3 control board and display
Troubleshooting
InjectMan
®
4
English (EN)
79
7
Troubleshooting
7.1
General errors
7.1.1
Motor module
7.1.2
Capillary
7.1.3
Control board and display
Symptom/message
Cause
Remedy
Motor axes move in the
wrong direction or do
not match joystick
movement.
• Motor module mounted
incorrectly.
• Incorrect installation
parameters entered.
• Direction of movement of
axis inverted.
Compare the installation
parameters with the
structure of the motor
module.
Check the orientation and
structure of the modules.
Check the connections of
the modules at the control
board.
Cancel the inversion of the
axis.
Symptom/message
Cause
Remedy
Capillary moves too
fast or too slow.
• Radius of working range
incorrectly set.
Set the radius using the
selection dial or in the
Speed
menu.
• Incorrect acceleration
factor.
In the
Installation
menu, set
the value for the
Dyn-factor
parameter.
Capillary only moves to
the side or vertically.
• Y-axis is disabled.
The
Y off
function disabled.
Capillary does not
move down far
enough.
• The
Z-axis Limit
function is
active.
• Capillary is incorrectly
adjusted.
Deactivate the
Z-axis Limit
function.
Readjust the capillary.
Symptom/message
Cause
Remedy
Device does not
respond to keystroke
when
Home
function is
active.
• Function is active.
Press the
home
key again.
The capillary moves down.
Press the
Back manual
softkey.
Move the joystick.