Troubleshooting

InjectMan

®

4

English (EN)

79

7

Troubleshooting

7.1

General errors

7.1.1

Motor module

7.1.2

Capillary

7.1.3

Control board and display

Symptom/message

Cause

Remedy

Motor axes move in the
wrong direction or do
not match joystick
movement.

• Motor module mounted

incorrectly.

• Incorrect installation

parameters entered.

• Direction of movement of

axis inverted.

Compare the installation
parameters with the
structure of the motor
module.

Check the orientation and
structure of the modules.

Check the connections of
the modules at the control
board.

Cancel the inversion of the
axis.

Symptom/message

Cause

Remedy

Capillary moves too
fast or too slow.

• Radius of working range

incorrectly set.

Set the radius using the
selection dial or in the

Speed

menu.

• Incorrect acceleration

factor.

In the

Installation

menu, set

the value for the

Dyn-factor

parameter.

Capillary only moves to
the side or vertically.

• Y-axis is disabled.

The

Y off

function disabled.

Capillary does not
move down far
enough.

• The

Z-axis Limit

function is

active.

• Capillary is incorrectly

adjusted.

Deactivate the

Z-axis Limit

function.

Readjust the capillary.

Symptom/message

Cause

Remedy

Device does not
respond to keystroke
when

Home

function is

active.

• Function is active.

Press the

home

key again.

The capillary moves down.

Press the

Back manual

softkey.

Move the joystick.