6.7.2

Moving to the position using the softkey

Prerequisites

• At least one position is stored.

1. Press a softkey with a stored position.

• You will move to the selected position.

• The joystick is deactivated until the position has been reached.

• The LEDs of the softkeys flash.

• In the softkey status field a filled circle is displayed.

6.7.3

Moving to the position using the joystick key

Prerequisites

• At least one position is stored.

• The

Joystick key

parameter must be set to the

Switch position

value.

6.7.4

Deleting a stored position

Prerequisites

• A position is stored.

1. When the position has been reached, hold the softkey.

• An acoustic signal will sound.

• The position has been deleted.

• The status field is empty

3. Move the capillary to the desired

position (e.g., parking position).

4. Press and hold the

Pos 2

softkey for

approx. one second to store the parking
position of the capillary.

• An acoustic signal will sound.

Pos 2

is marked.

• The stored position is displayed in the

status field.

If a stored position is lower than the lower limit (

Z-axis Limit

), the position on the

Z-axis will only be approached up to the defined limit.

Press the joystick key.
You will move to the first position

Press the joystick key.

• You will move to the next position.