2 execute the selftest function, 21 resetting parameters to the factory settings, 1 perform reset – Eppendorf InjectMan 4 User Manual
Page 78: 2 carrying out a reset in the menu, 22 remote computer control of the injectman 4
Operation
InjectMan
®
4
English (EN)
78
6.20.2
Execute the
Selftest
function
1. Select
Selftest
and confirm with
Enter
.
2. Confirm
Execute
with
Enter
.
The screen for the
Selftest
function appears.
3. Execute
Selftest
with
Start
.
The motors move up to the limit stops.
4. Terminate
Selftest
with
Stop
.
6.21
Resetting parameters to the factory settings
6.21.1
Perform reset
Reset all parameters to the factory settings.
1. Keep the
home
key pressed.
2. Switch on the device at the mains power switch.
The screen for
GENERAL RESET
appears.
3. Execute the function with
Yes
.
• All parameters are now reset to the factory settings.
6.21.2
Carrying out a reset in the menu
Prerequisites
• An application has been selected.
1. Press the
menu
key.
2. Select the
Function
menu.
3. Press
Enter
.
4. Select
User default
and confirm with
Execute
.
• All parameters are now reset to the factory settings.
6.22
Remote computer control of the InjectMan 4
You can control the InjectMan 4 by means of a terminal program remotely via computer.
Computer control by means of a terminal program is only recommended for certain
applications and is not described in this operating manual. A special operating manual for
this functionality is available at the following Internet address: www.eppendorf.com.
CAUTION! Risk of crushing between the modules
The modules automatically move in all spatial axes.
Do not reach into the movement range of the modules.
Eppendorf does not provide any support for controlling the InjectMan 4 by
means of a terminal program.