Page 73
Operation
InjectMan
®
4
English (EN)
73
6.17.2
Executing
Zero coordin.
1. Select the parameters and confirm with
Enter
.
2. Confirm
Execute
with
Enter
.
• All coordinates are reset to zero.
• Saved positions are deleted.
• Set limits are deleted.
6.17.3
Executing
Center motors
Prerequisites
• No capillary holder is clamped.
1. Select the parameters and confirm with
Enter
.
2. Confirm
Execute
with
Enter
.
• X-motor and Y-motor are moved to the middle position.
• Z-motor is moved to a 20/80 position.
• All coordinates are reset to zero.
• Saved positions are deleted.
• Set limits are deleted.
6.17.4
Executing
User default
1. Select the parameters and confirm with
Enter
.
2. Confirm
Execute
with
Enter
.
• All parameters are reset to the factory settings.
• The application screen appears.
Center motors
Move step motors of the motor module to the middle
position
Execute
User default
Reset the settings to the delivery status
Execute
CAUTION! Risk of crushing between the modules
The modules automatically move in all spatial axes.
Do not reach into the movement range of the modules.
Parameter
Value
Range of values