9.3
Test conditions
The errors specified apply only when the epT.I.P.S. pipette tips. Test conditions and test
evaluation in compliance with ISO 8655, Part 6. Test using a standardized fine balance with a
moisture trap.
•
Check the volumes in the Pip mode; keep the rocker pressed during the blow-out (Blow)
•
Set speed level: 5
•
Number of determinations: 10
•
Use of water in accordance with ISO 3696
•
Test at 20°C - 25°C ± 0.5°C
•
Dispensing against the tube wall
Rechargeable battery
Type:
Lithium polymer battery
Capacity:
750 mAh/3.7 V
Charging time:
Approx. 3 hours
Number of dispensing
operations:
Approx. 1400 in the Pip mode with fully charged battery and
average dispensing speed with a single-channel version
1 000 μL in continuous operation
Weight:
Approx. 20 g
Power supply
Type:
Power supply for Xplorer, Multipette/Repeater stream/
Xstream with plug-in power plug adapters
Input voltage:
100 V to 240 V ±10% 50/60 Hz
Output voltage:
5 V, 1 A
Xplorer
Operating temperature:
+5 to +40°C
Air humidity during operation:
10 to 95% rel. humidity
Storage temperature:
-5 to +45°C
Air humidity during storage:
10 to 95% rel. humidity
Weight without rechargeable
battery, without pipette tip:
100 μL single-channel approx.: 135 g (0.30 lb)
100 μL eight-channel approx.: 213 g (0.47 lb)
9